Continuous measurements may result in purpura, neuralgia and lack of blood. 
 
Do NOT wrap the cuff on limbs with transfusion tubes, intubations or skin lesions on the area, as damage may be 
caused to the limbs. 
 
The equipment can be used on pregnant or pre-eclamptic patients, but should not be used on neonatal patients. 
 
The operating steps needed to obtain accurate routine resting BLOOD PRESSURE measurements for the 
condition hypertension including: 
-- Patient position in normal use, including comfortably seated, legs uncrossed, feet flat on the floor, back and arm 
supported, middle of the cuff at the level of the right atrium of the heart. 
--The patient should be relax as much as possible and should not talk during the measurement procedure. 
-- 5 minutes should elapse before the first reading is taken. 
-- Operator position in normal use. 
 
Readings can be affected by the measurement site, the position of the patient  (standing, sitting, lying down), 
exercise, or the patient’s physiological condition. 
 
The operator should check whether the cuff is wrapped well at first if unexpected readings are obtained.