Bionime Corporation GM232 Manuale Utente

Pagina di 43
15
14
The GE006 Blood Glucose Test Strip
The GE006 Blood Glucose Test Strip
The GE006 Blood Glucose Meter can only be used with GE006 Blood Glucose Test Strips and Rightest™ 
Control Solution GC550. The use of other test strips or control solutions can lead to incorrect results.
Sample Port
Apply blood/control solution here. 
Noble Metal Electrodes
Electrochemical sensor
View WIndow
PRECAUTION
- Close the GE006 Blood Glucose Test Strip vial immediately after removing a test strip.
- Do not reuse GE006 Blood Glucose Test Strips.
- Do not use expired GE006 Blood Glucose Test Strips. 
- When you open a new vial of GE006 Blood Glucose Test Strips, record the date on the vial. 
Discard the vial of test strips after 3 months from opening.
- Store the GE006 Blood Glucose Test Strip between 4-30°C (or 39-86°F)  and in a location (< 
90% relative humidity). Do not expose to direct sunlight or heat.
- If the GE006 Blood Glucose Meter and Test Strips are exposed to a substantial change in 
temperature, please wait 45 minutes before measurement.
- For detailed information, please refer to the GE006 Blood Glucose Test Strip Package 
Insert.
Electrode Contacts
Hold here to insert 
test strip into 
meter.