Bionime Corporation GM782 Manuale Utente

Pagina di 32
51
50
Er1 - 
Er2 - The m
Er3 - The s
The inserted test strip has been used or damaged. Please use a new test 
strip.
eter has malfunctioned. Do a Quality Control Test or reinstall the 
batteries to check if the meter works properly.
ignal transmission is disrupted. Repeat the test using a new test strip.
Repeat the test using a new test 
strip.
Er5 - An issue calibrating the Meter has occurred. Please follow the steps below:
  (1) Remove the test strip from the meter.
  (2) Turn off the meter (press and hold the Main button for 3 seconds).
  (3) Press the main button to turn ON the meter.
  If after turning on the Meter, you do not see Er5, your meter is functioning 
properly and able to perform a test.
If one of the above error messages still appears, please contact Bionime  
Customer Service.
Er4 - The blood sample volume is insufficient. 
Error Messages and Troubleshooting
Meter Malfunction
If the 
 Blood Glucose Meter
 will not turn on, please follow the steps below:
1. Open the battery cover and remove the batteries.
2. Wait for 5 minutes and reinsert the batteries as described as instructed on page 16, (Meter 
Activation and Changing the Battery).
The meter should work normally after finishing the above steps. If the meter still does not work, 
please contact your local 
 customer service.
TM
 Rightest
GM782
 
Bionime
Strip Error
"           "
TM
Rightest
"          "
"          "
TM
Rightest
TM
Rightest
Sampling Error
Please do not apply the blood to the sample port of the test strip before the 
meter displays "      ". Discard the test strip If the meter shows "          ".
1. When the test strip is inserted incorrectly, the meter will beep 4 times while the 
 symbol flashes. Do not apply a sample on the test strip. Please re-
insert the unused test strip correctly (refer to “Handling the 
 Blood 
Glucose Test Strip GS700“ on Page 23 for more information).
2. If, after performing the required steps, the 
 and flashing 
 
symbols appear on the screen, it means the 
 Blood Glucose Test 
Strip GS700 has been inserted incorrectly more than twice. Please, re-insert 
the unused test strip again after reviewing the steps on how to insert a test 
strip correctly on p.23 or call Customer Service for support. This error message 
can also indicate that you may have used the wrong test strips. Please check 
the test strip vial to ensure you are using the 
 Blood Glucose Test 
Strip GS700. 
Battery Error
1. The "         " and "          " symbols appear and the meter will not function when 
the battery is low. Change the batteries immediately.
Error Messages and Troubleshooting