Limitations
Error Messages and Troubleshooting
 Blood Glucose Monitoring System GM782 is not intended for serum or plasma testing.
- Severe dehydration and excessive water loss may cause inaccurately low results.
- The glucose test may be invalid in the presence of abnormally high concentration: Ascorbic acid 
( 6 mg/dL), Dopamine (  2.5 mg/dL), 
  L-Dopa (  3 mg/dL), 
.
- Not for screening or diagnosis of diabetes mellitus.
- Not for use on critically ill patients in shock, dehydrated patients or hypo-osmolar patients. 
- Alternative site testing (AST) should only be performed during steady-state times (when glucose is 
not changing rapidly).
- Alternative Site Testing should not be used to calibrate continuous glucose monitoring systems 
(CGMS).

- Results from alternative site testing should not be used in insulin dose calculation.
- Regarding potential electromagnetic or other interference, keep meter out of electromagnetic 
radiation sources.
TM
Rightest
- Inaccurate test results may be obtained at altitudes greater than 10,000 feet (3048 meters) above 
sea level.
 
Xylose (≧ 20 mg/dL) and Uric Acid (≧ 20 mg/dL)
Alternative site testing should not be performed if testing for hypoglycemia (Low blood glucose).
CAUTION
A blood sample should ONLY be applied to the test strip after the test strip has been 
inserted correctly and the Meter's screen is showing the image of the test strip and a blood 
drop flashing symbol. If the blood drop flashing symbol is NOT showing on the screen, do 
not apply a sample to the test strip. Please re-insert the unused test strip correctly. The 
screen MUST show the image of the test strip accompanied by the blood drop symbol 
flashing (this symbol take 3 seconds to show after correctly inserting the test strip) in order 
for a sample to be applied. Please consult your User Manual and/or contact Customer 
Service for support on how to correctly insert a 
 Blood Glucose Test Strip GS700.