NIBP: NIBP modes of operation
NIBP modes of operation
The monitor has several NIBP modes that can be
selected by the user. NIBP determinations are
automated and, upon completion, the values for
systolic pressure, diastolic pressure, mean arterial
pressure, and pulse rate (if SpO
selected by the user. NIBP determinations are
automated and, upon completion, the values for
systolic pressure, diastolic pressure, mean arterial
pressure, and pulse rate (if SpO
2
is not active) are
shown in their respective windows. NIBP can be
measured once or have as many measurements as
possible within a given time frame using STAT, Cycle
or Profile. NIBP auto mode can be selected in
Monitor Setup > NIBP or selecting the auto mode icon on the screen and then
making the selection.
Profile mode can be configured in password-protected Default Setup. The
Profile settings allow the user control over the sequence of these
determinations. The user may establish multiple groups of NIBP determination
cycle sequences, each with a defined number of NIBP intervals with a defined
duration. When that sequence expires, the next sequence takes over until all
sequences have completed. As an example; one Profile could comprise of 4
determinations at 5-minute intervals followed by 8 determinations at 60-minute
intervals, resulting in a total of 12 determinations over a period of 8 hours and 15
minutes.
measured once or have as many measurements as
possible within a given time frame using STAT, Cycle
or Profile. NIBP auto mode can be selected in
Monitor Setup > NIBP or selecting the auto mode icon on the screen and then
making the selection.
Profile mode can be configured in password-protected Default Setup. The
Profile settings allow the user control over the sequence of these
determinations. The user may establish multiple groups of NIBP determination
cycle sequences, each with a defined number of NIBP intervals with a defined
duration. When that sequence expires, the next sequence takes over until all
sequences have completed. As an example; one Profile could comprise of 4
determinations at 5-minute intervals followed by 8 determinations at 60-minute
intervals, resulting in a total of 12 determinations over a period of 8 hours and 15
minutes.
NOTES
If an alarm condition emerges and the silence icon is selected, the current
determination will be completed, but subsequent repeated determinations
will not be carried out.
determination will be completed, but subsequent repeated determinations
will not be carried out.
When an automated STAT, Cycle or Profile measurement is being
performed, the auto mode button in the Monitor Setup > NIBP is disabled.
performed, the auto mode button in the Monitor Setup > NIBP is disabled.
Adaptive target inflation pressure
If several NIBP determinations are performed on the same patient, the monitor
will adjust the target inflation pressure based on previous determination results
for that patient. Adaptive target inflation pressure is always patient-specific.
When the patient is discharged, results of his/her NIBP determinations will not
be used for the next patient. Adaptive history is cleared if a single NIBP
measurement is not performed for more than two minutes. Adaptive history is
not cleared in STAT, Cycle or Profile modes.
will adjust the target inflation pressure based on previous determination results
for that patient. Adaptive target inflation pressure is always patient-specific.
When the patient is discharged, results of his/her NIBP determinations will not
be used for the next patient. Adaptive history is cleared if a single NIBP
measurement is not performed for more than two minutes. Adaptive history is
not cleared in STAT, Cycle or Profile modes.