NIBP: NIBP modes of operation
NIBP modes of operation
The monitor has several NIBP modes that can be 
selected by the user. NIBP determinations are 
automated and, upon completion, the values for 
systolic pressure, diastolic pressure, mean arterial 
pressure, and pulse rate (if SpO
2
 is not active) are 
shown in their respective windows. NIBP can be 
measured once or have as many measurements as 
possible within a given time frame using STAT, Cycle 
or Profile. NIBP auto mode can be selected in 
Monitor Setup > NIBP or selecting the auto mode icon on the screen and then 
making the selection.
Profile mode can be configured in password-protected Default Setup. The 
Profile settings allow the user control over the sequence of these 
determinations. The user may establish multiple groups of NIBP determination 
cycle sequences, each with a defined number of NIBP intervals with a defined 
duration. When that sequence expires, the next sequence takes over until all 
sequences have completed. As an example; one Profile could comprise of 4 
determinations at 5-minute intervals followed by 8 determinations at 60-minute 
intervals, resulting in a total of 12 determinations over a period of 8 hours and 15 
minutes.
NOTES
If an alarm condition emerges and the silence icon is selected, the current 
determination will be completed, but subsequent repeated determinations 
will not be carried out.
When an automated STATCycle or Profile measurement is being 
performed, the auto mode button in the Monitor Setup > NIBP is disabled.
Adaptive target inflation pressure
If several NIBP determinations are performed on the same patient, the monitor 
will adjust the target inflation pressure based on previous determination results 
for that patient. Adaptive target inflation pressure is always patient-specific. 
When the patient is discharged, results of his/her NIBP determinations will not 
be used for the next patient. Adaptive history is cleared if a single NIBP 
measurement is not performed for more than two minutes. Adaptive history is 
not cleared in STATCycle or Profile modes.