3 Specifications 
8 
Feature  
Values  
Resolution 
Temperature: 0.1 °C / 0.1°F 
Humidity: 0.1% RH 
Flow velocity: 0.01 m/s 
Volume flow: 1 m³/h / 1 cfm 
Pressure (absolute pressure): 0.1 hPa / 
0.1 mbar / 0.01 kPa 
Pressure (differential pressure): 0.001 Pa / 
0.00001 hPa / 0.00001 mbar / 0.0001 mm 
H2O / 0.000001 inH2O 
Accuracy 
(nominal temperatur
e 22°C/71.6°F) 
Temperature: ±0.5 °C (0 to +70 °C) / ±0.8 °C 
(-20 to 0 °C) 
 
Humidity: ±1.8% RH +3% of m.v. at +22 °C 
(5 to 80% RH) (longer high humidity application 
can result in a temporary sensor drift) 
 
Flow velocity: no accuracy specification since 
calculated variable 
 
Volume flow 
1
: ±3% of m.v. +12 m³/h at 
+22 °C, 1013 hPa (85 to 3500 m³/h) 
±3% +7 cfm (50 to 2,100 cfm) 
Error absolute pressure compensation: 
±0.04% of m.v. / hPa deviating from 1013 hPa 
 
Pressure (absolute pressure): ± 5 hPa 
 
Pressure (differential pressure): ±2% of m.v. 
+ 0.5 Pa (at 22°C, 1013 hPa) 
Error absolute pressure compensation: 
±0.04% of m.v. / hPa deviating from 1013 hPa 
                                                           
1
 All accuracy specifications apply under laboratory conditions or with necessary 
compensation (correction factor) with the standard hood 610x610 mm.