testo Instruments Co. Ltd 05600420 Manuale Utente

Pagina di 24
3 Specifications 
 
 
9 
Feature  
Values  
Temperature 
coefficient 
Humidity: ± 0.03% RH / K (deviating from 
22°C, in the range 0 to 60 °C) 
 
Volume flow: ± 0.02% of m.v. / K 
(deviating from 22 °C, in the range 0 to 60 °C) 
 
Pressure (absolute pressure): ± 0.02% of 
m.v. / K (deviating from 22 °C, in the range 
0 to 60 °C) 
 
Pressure (differential pressure): ± 0.02% of 
m.v. / K (deviating from 22 °C, in the range 
0 to 60 °C) 
Response time t90  Temperature: approx. 45 s 
Humidity: approx. 15 s 
Flow velocity: approx. 1 s 
Volume flow: approx. 1 s 
Pressure (absolute pressure): approx. 1 s 
Pressure (differential pressure): approx. 1 s 
Operating and 
ambient conditions 
Storage temperature: -20 to +60 °C / -4 to 
140 °F 
Operating temperature: -5 to +50 °C/-4 to 
122 °F 
Humidity: 0 to 100% RH  
Pressure range: 800 to 1100 hPa 
Housing / 
measurement setup 
Measuring instrument housing material: ABS 
Body material: PP 
Standard hood material: Nylon 
Measuring instrument dimensions: 
150x85x35 mm 
Body dimensions: 490x970x610 mm 
Dimensions of measurement setup with 
standard hood: 610x970x610 mm 
Weight of entire measurement setup 
approx. 2900 g  
Power supply 
4 x 1.5 V rechargeable/non-rechargeable 
batteries  
Type AA / alkaline manganese, mignon 
Battery lifetime: approx. 40h 
(display illumination off, Bluetooth off)