Save - Press the Save option button to store your mask. 
After pressing, the mask is saved Internally, or the SAVE 
ALL
 screen is shown, depending on your Location control 
setting (choose Internal or External as desired on Mask - 
Page 1/2). 
 
Load - Press the Load option button to retrieve a stored 
mask. After pressing, the mask is retrieved Internally, or the 
SAVE ALL screen is shown, depending on your Location 
control setting (choose Internal or External as desired on 
Mask - Page 1/2). 
 
Return - Press the Return option and the Pass/Fail menu is 
shown. 
4.  Press the Return option button or the Utility front panel button 
and the Utility menu is shown. 
Using Record 
The Record function can record the input waveform from CH1 and CH2 
(also, CH3 and CH4 on four channel versions), with a maximum record 
length of 2500 frames. This keeps you from having to watch the signal 
when capturing abnormal signals over a long period of time. 
Access the Record function by pressing the corresponding option button 
on the Utility menu. The Record menus are then shown.