Lecroy WAVEACE 1012 2-channel oscilloscope, Digital Storage oscilloscope, WaveAce 1012 Manuale Utente

Codici prodotto
WaveAce 1012
Pagina di 123
Operator's Manual 
40 
WA1K2K-OM-E RevB 
Pulse Triggering 
Use Pulse width triggering to trigger on aberrant pulses. You can select 
how to compare the trigger pulse relative to the pulse width as follows: 
 
 Positive pulse width less than pulse width setting 
 
 Positive pulse width larger than pulse width setting 
 
 Positive pulse width equal to pulse width setting 
 
 Negative pulse width less than pulse width setting 
 
 Negative pulse width larger than pulse width setting 
 
 Negative pulse width equal to pulse width setting 
1.  Press the Type option button and select Pulse
 
2.  Press the Source option button to choose a channel input (CH1 - 
2; or CH1 - 4 on 4 channel models) or EXTEXT/5AC Line input.   
3.  Press the When option button to select how to compare the 
trigger pulse relative to the value selected in the Set Width option. 
4.  Press the Set Width option button and then turn the Adjust front 
panel knob to set the pulse width. 
5.  Press the Next Page option button. 
6.  Press the Mode option button to select AutoNormal, or Single 
mode. Use Auto mode to let the acquisition automatically run in 
the absence of a trigger. Use Normal mode when you want to see 
only valid triggered waveforms (when you use this mode the 
oscilloscope does not display a waveform until after the first