the absence of a trigger. Use Normal mode when you want to see 
only valid triggered waveforms (when you use this mode the 
oscilloscope does not display a waveform until after the first 
trigger). Use Single mode when you want the oscilloscope to 
acquire a single waveform.  
8.  Press the Setup... option button to display the Trigger Setup... 
menu. You can use the Trigger Setup... menu to select a coupling 
mode and define a Holdoff value. Coupling modes comprise DC
ACHF Reject, and LF Reject. Turn the Adjust front panel knob to 
set the Holdoff value. 
Slope Triggering 
Use Slope Triggering to trigger on the positive slope or negative slope 
depending on the trigger conditions and time you set. 
1.   Press the Type option button and select Slope
 
2.  Press the Source option button to choose a channel input (CH1 - 
2; or CH1 - 4 on 4 channel models) or EXTEXT/5AC Line input. 
3.  Press the When option button to select the trigger condition. 
4.  Press the Time option button and then turn the Adjust front panel 
knob to set the slope time. 
5.  Press the Next Page option button. 
6.  Press the Vertical option button and select the trigger level that 
can be adjusted using the Level front panel knob. You can adjust 
Level A, Level B, or adjust both at the same time.