Lecroy WAVEACE 2022 2-channel oscilloscope, Digital Storage oscilloscope, WaveAce 2022 Manuale Utente
Codici prodotto
WaveAce 2022
WaveAce Remote Control
26
920836 Rev A
ACQUISITION - TRIG_LEVEL, TRLV
Description
The TRIG_LEVEL command adjusts the trigger level of the specified trigger source. An out-of-
range value will be adjusted to the closest legal value.
range value will be adjusted to the closest legal value.
The TRIG_LEVEL? query returns the current trigger level.
Command Syntax
[<trig_source>]:TRig_LeVel <trig_level>
<trig_source>:= {C1,C2,C3,C4,EX5}
Query Syntax
<trig_source>:TRig_LeVel?
Response Format
<trig_source>:TRig_LeVel <trig_level>
Example
The following instruction adjusts the trigger level of Channel 2 to -3.4 V:
C2:TRLV -3.4V
Related Commands
TRIG_COUPLING, TRIG_DELAY, TRIG_MODE, TRIG_SELECT, TRIG_SLOPE