ACQUISITION - TRIG_SELECT, TRSE 
Description 
The TRIG_SELECT command selects the condition that triggers the acquisition of waveforms. 
Only the Edge trigger is supported. 
Command Syntax 
TRig_Select EDGE,SR,<source>,HT,TI,HV,<hold_value>  
<source>:= {C1,C2,C3,C4,LINE,EX,EX5} 
<hold_value>:=
  Refer to the trigger information provided for your oscilloscope model. 
NOTE: When specifying <hold_value>, the unit S (seconds) is optional. 
Query Syntax 
TRig_SElect? 
Response Format 
TRig_SElect <trig_type>,SR,<source>,HT,<hold_type>,HV,<hold_value> 
HV2 only returned if <hold_type> is P2 or I2. 
Example 
The following instruction selects the single-source trigger with Channel 1 as trigger source.  
Hold type and hold value are chosen as "pulse smaller" than 20 ns: 
TRSE SNG,SR,C1,HT,PS,HV,20 NS