Gw Instek GPT-9802 Insulation measuring device, GPT-9802 Manuale Utente

Codici prodotto
GPT-9802
Pagina di 156
 
 
GPT-9000 Series User Manual 
92
 
Exit Testing 
To exit testing, press the 
MANU/AUTO key when the 
tester is in the READY status. The 
tester will revert to the VIEW 
status for the current automatic 
test. 
MANU/AUTO
 
 
I R  
G B  
A CW
D CW
 
 
E        
 
 
F R E Q =  
0 H z  
6
0
   
100
k
 
V
 
       
R
A C
 
=
F F
O
E
I
V
 
 
                         
M
T I
E
= 0 0 1 . 0 S
R
0 0 m A
E
R F
=
#
0 .
0
1 . 0 0 m A
I
H
E
S
= 0
T
0 0 . 1 S
0
A M P
=
R
 
W
M
A U O
T
N A
_
0
A U T O = 0 0 1 - 1 0
 
 
m A
 
Automatic Test Results 
 
Background 
If all the test steps are allowed to run to 
completion (the AUTO test is not stopped or a 
protection setting is not tripped) then the tester 
will judge each step as either PASS or FAIL. 
This is shown as a table after the automatic test 
has finished running. 
Overview 
P A S
P A S
S
S
P A S
S K I
S
P
E        
 
M
A U O
T
N A
_
 
1
A U T O = 0 0 1 - 0 0
# 0 1 :
# 0 2 :
# 0 3 :
# 0 4 : P A S
# 0 5 :
# 0 6 :
# 0 7 :
# 0 8 : S T O
# 0 9 :      
# 1 0 :      
# 1 1 :      
# 1 2 :      
# 1 3 :      
# 1 4 :      
# 1 5 :      
# 1 6 :      
S
P
F A I
F A I
L
L
FAIL judgment
PASS judgment
skipped step
step stopped  
 Note 
The PASS/FAIL judgment for an automatic test 
depends on the results of all the steps (manual 
tests) that compose the automatic test: 
 
Each step must be passed for a PASS judgment 
(excluding skipped tests). 
 
A FAIL result for a single step will result in 
FAIL for the whole automatic test.