GT-600/GT-800 Instruction Manual
65
I
PE(
)
-
I
SUB INT
0.15 mA
Path pre-test (self-test).
If the instrument’s measurement path is interrupted e.g.
due to blown internal fuse, then “ERROR 1” message
will be displayed and further I
If the instrument’s measurement path is interrupted e.g.
due to blown internal fuse, then “ERROR 1” message
will be displayed and further I
PE(
)
tests will be disabled.
-
I
SUB
20.00 mA
Pre-test.
If the substitute leakage current exceeds 20.00 mA
(damaged appliance), then “LN-PE SHORTED!” message
will be displayed and further I
If the substitute leakage current exceeds 20.00 mA
(damaged appliance), then “LN-PE SHORTED!” message
will be displayed and further I
PE(
)
tests will be disabled.
-
I
L
25 mA
Pre-test.
If the load current is lower than 25 mA (open mains ON
switch), then “NO DEVICE!” message will be displayed
and further I
If the load current is lower than 25 mA (open mains ON
switch), then “NO DEVICE!” message will be displayed
and further I
PE(
)
tests will be disabled. The operator may
ignore the message by pressing the IGNR key.
-
Grounding of mains
plug PE terminal
Pre-test.
If the PE terminal is not sufficiently grounded, then “PE
DISCONNECTED! TURN OFF NOW!” message will be
displayed and any further operation will be disabled.
Switch off the tester immediately, danger!
If the PE terminal is not sufficiently grounded, then “PE
DISCONNECTED! TURN OFF NOW!” message will be
displayed and any further operation will be disabled.
Switch off the tester immediately, danger!
-
Fuse F2
Pre-test.
If the F2 fuse is blown, then “FUSE 2!” message will be
displayed and further I
If the F2 fuse is blown, then “FUSE 2!” message will be
displayed and further I
PE(
)
tests will be disabled.
-
U
EXT PROBE1/GND
50 V
-
U
EXT PROBE2/GND
50 V
Pre-test.
If an external voltage higher than 50 V is detected, then
“EXTERNAL VOLTAGE!” message will be displayed and
further I
If an external voltage higher than 50 V is detected, then
“EXTERNAL VOLTAGE!” message will be displayed and
further I
PE(
)
tests will be disabled.
I
T
-
I
SUB INT
0.15 mA
Path pre-test (self-test).
If the instrument’s measurement path is interrupted e.g.
due to blown internal fuse, then “ERROR 1” message
will be displayed and further I
If the instrument’s measurement path is interrupted e.g.
due to blown internal fuse, then “ERROR 1” message
will be displayed and further I
T
tests will be disabled.
-
I
SUB
3.50 mA
Pre-test.
If the substitute leakage current exceeds 3.50 mA
(damaged appliance), then “LN-PE SHORTED!” message
will be displayed and further I
If the substitute leakage current exceeds 3.50 mA
(damaged appliance), then “LN-PE SHORTED!” message
will be displayed and further I
T
tests will be disabled.
-
I
L
25 mA
Pre-test.
If the load current is lower than 25 mA (open mains ON
switch), then “NO DEVICE!” message will be displayed
and further I
If the load current is lower than 25 mA (open mains ON
switch), then “NO DEVICE!” message will be displayed
and further I
T
tests will be disabled. The operator may
ignore the message by pressing the IGNR key.
-
Grounding of mains
plug PE terminal
Pre-test.
If the PE terminal is not sufficiently grounded, then “PE
DISCONNECTED! TURN OFF NOW!” message will be
displayed and any further operation will be disabled.
Switch off the tester immediately, danger!
If the PE terminal is not sufficiently grounded, then “PE
DISCONNECTED! TURN OFF NOW!” message will be
displayed and any further operation will be disabled.
Switch off the tester immediately, danger!
-
Fuse F2
Pre-test.
If the F2 fuse is blown, then “FUSE 2!” message will be
displayed and further tests will be disabled.
If the F2 fuse is blown, then “FUSE 2!” message will be
displayed and further tests will be disabled.
P/I
L
-
I
SUB INT
0.15 mA
Path pre-test (self-test).
If the instrument’s measurement path is interrupted e.g.
due to blown internal fuse, then “ERROR 1” message
will be displayed and further P/I
If the instrument’s measurement path is interrupted e.g.
due to blown internal fuse, then “ERROR 1” message
will be displayed and further P/I
L
tests will be disabled.
-
I
SUB
3.50 mA
Pre-test.
If the substitute leakage current exceeds 3.50 mA
(damaged appliance), then “LN-PE SHORTED!” message
will be displayed and further P/I
If the substitute leakage current exceeds 3.50 mA
(damaged appliance), then “LN-PE SHORTED!” message
will be displayed and further P/I
L
tests will be disabled.
-
Grounding of mains
plug PE terminal
Pre-test.
If the PE terminal is not sufficiently grounded, then “PE
DISCONNECTED! TURN OFF NOW!” message will be
displayed and any further operation will be disabled.
Switch off the tester immediately, danger!
If the PE terminal is not sufficiently grounded, then “PE
DISCONNECTED! TURN OFF NOW!” message will be
displayed and any further operation will be disabled.
Switch off the tester immediately, danger!