GT-600/GT-800 Instruction Manual 
65 
 
 
I
PE(
)
 
-
 I
SUB INT
 
0.15 mA 
Path pre-test (self-test).
If the instrument’s measurement path is interrupted e.g. 
due to blown internal fuse, then “ERROR 1” message 
will be displayed and further I
PE(
) 
tests will be disabled. 
-
 I
SUB
 
20.00 mA 
Pre-test.
If the substitute leakage current exceeds 20.00 mA 
(damaged appliance), then “LN-PE SHORTED!” message 
will be displayed and further I
PE(
) 
tests will be disabled. 
-
 I
25 mA 
Pre-test.
If the load current is lower than 25 mA (open mains ON 
switch), then “NO DEVICE!” message will be displayed 
and further I
PE(
) 
tests will be disabled. The operator may 
ignore the message by pressing the IGNR key. 
-
 Grounding of mains 
plug PE terminal 
Pre-test.
If the PE terminal is not sufficiently grounded, then “PE 
DISCONNECTED! TURN OFF NOW!” message will be 
displayed and any further operation will be disabled. 
Switch off the tester immediately, danger! 
-
 Fuse F2
Pre-test.
If the F2 fuse is blown, then “FUSE 2!” message will be 
displayed and further I
PE(
) 
tests will be disabled. 
-
 U
EXT PROBE1/GND 
50 V 
-
 U
EXT PROBE2/GND 
50 V 
Pre-test.
If an external voltage higher than 50 V is detected, then 
“EXTERNAL VOLTAGE!” message will be displayed and 
further I
PE(
) 
tests will be disabled. 
I
T
 
-
 I
SUB INT
 
0.15 mA 
Path pre-test (self-test).
If the instrument’s measurement path is interrupted e.g. 
due to blown internal fuse, then “ERROR 1” message 
will be displayed and further I
T
 tests will be disabled. 
-
 I
SUB
 
3.50 mA 
Pre-test.
If the substitute leakage current exceeds 3.50 mA 
(damaged appliance), then “LN-PE SHORTED!” message 
will be displayed and further I
T
 tests will be disabled. 
-
 I
25 mA 
Pre-test.
If the load current is lower than 25 mA (open mains ON 
switch), then “NO DEVICE!” message will be displayed 
and further I
T
 tests will be disabled. The operator may 
ignore the message by pressing the IGNR key. 
-
 Grounding of mains 
plug PE terminal 
Pre-test.
If the PE terminal is not sufficiently grounded, then “PE 
DISCONNECTED! TURN OFF NOW!” message will be 
displayed and any further operation will be disabled. 
Switch off the tester immediately, danger! 
-
 Fuse F2
Pre-test.
If the F2 fuse is blown, then “FUSE 2!” message will be 
displayed and further tests will be disabled. 
P/I
L
 
-
 I
SUB INT
 
0.15 mA 
Path pre-test (self-test).
If the instrument’s measurement path is interrupted e.g. 
due to blown internal fuse, then “ERROR 1” message 
will be displayed and further P/I
L
 tests will be disabled. 
-
 I
SUB
 
3.50 mA 
Pre-test.
If the substitute leakage current exceeds 3.50 mA 
(damaged appliance), then “LN-PE SHORTED!” message 
will be displayed and further P/I
L
 tests will be disabled. 
-
 Grounding of mains 
plug PE terminal 
Pre-test.
If the PE terminal is not sufficiently grounded, then “PE 
DISCONNECTED! TURN OFF NOW!” message will be 
displayed and any further operation will be disabled. 
Switch off the tester immediately, danger!