Beha Amprobe GT-800 STD KITVDE-tester 4472062 Manuale Utente

Codici prodotto
4472062
Pagina di 76
GT-600/GT-800 Instruction Manual 
66 
 
 
 
-
 Fuse F2
Pre-test.
If the F2 fuse is blown, then “FUSE 2!” message will be 
displayed and further P/I
L
 tests will be disabled. 
CLAMP -
 Internal fuse 
If internal fuse is blown, then “ERROR 2!” message will 
be displayed and further CLAMP tests will be disabled. 
CORD 
-
 U
L IEC
 
50 V 
-
 U
N IEC
 
50 V 
Pre-test.
If an external voltage higher than 50 V is detected after 
pressing START key, then “EXTERNAL VOLTAGE!” 
message will be displayed and further CORD tests will be 
disabled. 
-
 U
EXT PROBE1/GND 
50 V 
-
 U
EXT PROBE2/GND 
50 V 
Pre-test and during the R
PE
test.
If an external voltage higher than 50 V is detected, then 
“EXTERNAL VOLTAGE!” message will be displayed and 
further CORD tests will be disabled. 
-
 Fuse F3
Pre-test.
If the F3 fuse is blown, then “FUSE 3!” message will be 
displayed and further CORD tests will be disabled. 
PRCD 
-
 U
L IEC
 
50 V 
-
 U
N IEC
 
50 V 
Pre-test and during the R
PE
test.
If an external voltage higher than 50 V is detected, then 
“EXTERNAL VOLTAGE!” message will be displayed and 
further PRCD tests will be disabled. 
-
 I
SUB INT
 
0.15 mA 
Path pre-test (self-test).
If the instrument’s measurement path is interrupted e.g. 
due to blown internal fuse, then “ERROR 1” message 
will be displayed and further P/I
L
 tests will be disabled. 
-
 Grounding of mains 
plug PE terminal 
Pre-test.
If the PE terminal is not sufficiently grounded, then “PE 
DISCONNECTED! TURN OFF NOW!” message will be 
displayed and any further operation will be disabled. 
Switch off the tester immediately, danger! 
-
 Fuse F2
Pre-test.
If the F2 fuse is blown, then “FUSE 2!” message will be 
displayed and further PRCD tests will be disabled. 
-
 I
SUB
 
3.50 mA 
Pre-test.
If the substitute leakage current exceeds 3.50 mA 
(damaged appliance), then “LN-PE SHORTED!” message 
will be displayed and further PRCD tests will be disabled. 
-
 I
SUB
 
20.00 mA 
Pre-test.
If the substitute leakage current exceeds 20.00 mA 
(damaged appliance), then “LN-PE SHORTED!” message 
will be displayed and further CORD
 
tests will be disabled. 
-
 Fuse F3
Pre-test in R
PE
measurement.
-
 U
EXT PROBE1/GND 
50 V  Pre-test and during the measurement in R
PE
 
measurement 
-
 U
EXT PROBE2/GND 
50 V  Pre-test and during the measurement in R
PE 
measurement