Beha Amprobe GT-800 STD KITVDE-tester 4472062 Manuale Utente
Codici prodotto
4472062
GT-600/GT-800 Instruction Manual
69
Accuracy, class I, II
± (5% rdg. + 3 digits) 0.00 ... 20.0 M
± (8% rdg.) 20.1 ... 50.0 M
± (15% rdg.) 51 ... 100 M
Operational error, class I, II
± 15%
Accuracy, class III
± (7% rdg. + 3 digits) 0.00 ... 20.0 M
Operational error, class III
± 14%
Limit value
Adjustable 0.01 ... 20.0 M
Standard limit values:
0.30
M
(class I with heater),
1.00
M
(class I),
2.00
M
(class II),
0.25
M
(class III)
Test voltage (U
N
)
500 V DC (and 250 V DC GT-800 only),
(floating
output)
Test voltage tolerance
-0% ... +25%
Test current (500 V range)
>1 mA DC (at 500 k
load)
Test current (250 V range)
>1 mA DC (at 250 k
load) (GT-800 only)
Short-circuit current
< 8 mA DC
Capacitive load (operating func.)1
F max. in parallel with R
INS
Discharge (class I and II)
Via internal resistance of 200 k
after finishing
the measurement
Discharge (class III)
Via internal resistance of 3 M
after finishing
the measurement
Pre-Tests and Protection,
Functions and Messages
Functions and Messages
Refer to page 64
Notes!
Class I/II: Do not ground LN terminals of test socket. The grounding will influence
test result with parallel connection of 3.3 M
approx. as a result of tester’s
internal resistance.
Class III: DUT must be isolated from ground, otherwise test result will not be
correct.
Class III: DUT must be isolated from ground, otherwise test result will not be
correct.
Substitute Leakage Current TRMS (I
SUB
)
Available protection class
I and II
Measurement terminals class I
L/N against (PE on test socket in parallel with
PROBE
2)
Measurement terminals class II L/N against PROBE 2
Measurement range
Measurement range
0.25 ... 20.00 mA
Display range
0.25 ... 20.00 mA
Resolution 0.01
mA
Accuracy
+/- (5% rdg. + 3 digits)
Operational error
± 12%
Limit value
Adjustable 0.25 ... 20.00 mA
Standard limit values:
3.50 mA (class I)
0.50 mA (class II)
Open-circuit test voltage
Approx. 45 V / 50 Hz at 230 V, SELV (floating
output)
Short-circuit test current
Short-circuit test current
Approx. 300 mA