135 
 
Table3 
 
 
Y Range constant- Range Hold 
 
Component 
 
Inductor 
Capacitor 
Resistor 
Range 
Max 
Min 
Max 
Min 
Max 
Min 
Range1 
16mH/f 
1mH/f 
25uF/f 
1.6uF/f 
100Ω 
6.25Ω 
Range2 
256mH/f  16mH/f  1600nF/f  100nF/f 
1.6kΩ 
0.1kΩ 
Range3 
4100mH/f  256mH/f  100nF/f 
6.4nF/f 
25.6kΩ 
1.6kΩ 
Range4* 
65H/f 
4.1H/f 
6400pF/f  400pF/f 
410kΩ 
25.6kΩ 
f= test frequency in kHz 
*  This range is not used above 20 kHz  
 
 
 
 
 
 
 
 
Table4 
 
 
Y Range constant- Auto Range 
 
Component 
 
Inductor 
Capacitor 
Resistor 
Range 
Max 
Min 
Max 
Min 
Max 
Min 
Auto range  65H/f** 
1mH/f 
25uF/f  400pF/f **  410kΩ**  6.25Ω** 
**: Above 20kHz, Cmin = 6.4 nF/f, and Lmax = 4100mH/f 
f = test frequency in kHz. 
   LCR-800 
User 
Manual 
136
 
Specifications 
Specification accuracy is only applicable when the LCR meter has 
been warmed up for 30 minutes with an operating temperature of 
18°C ~28°C.  
 
Measurement 
Parameters 
Inductance (Ls/Lp)*, Capacitance (Cs/Cp), Resistance 
(Rs/Rp), Dissipation (D), Quality Factors (Q), 
Equivalent Series Resistance (ESR) and Equivalent 
Parallel Resistance (EPR), Impedance (|Z|), Phase 
angle of Impedance [degree] (θ). 
Measurement 
Modes 
R/Q, C/D, C/R, L/Q, Z/
 
θ, L/R 
Display Ranges 
Primary 
Display 
Inductance 
(L) 
0.00001mH ~ 99999H 
  
Capacitance 
(C) 
0.00001pF ~ 99999μF 
 
 
Resistance 
(R) 
0.00001Ω ~ 99999kΩ 
  
Absolute 
of 
Impedance 
(Z) 
0.00001Ω ~ 99999kΩ 
 
Secondary 
Display 
Dissipation 
factor (D)
+
 
0.0001 ~ 9999 
  
Quality 
factor (Q)** 
0.0001 ~ 9999 
 
 
Phase angle 
of 
Impedance 
(degree) 
-180.00∘ ~ 180.00∘ 
  
Equivalent 
Series 
Resistance 
(ESR)
+
 
0.0001Ω ~ 9999 kΩ 
 
 
Equivalent 
Parallel 
Resistance 
(EPR)
+
 
0.0001Ω ~ 9999 kΩ