Gw Instek LCR-816 Digital-Multimeter, DMM, LCR-816 Scheda Tecnica

Codici prodotto
LCR-816
Pagina di 91
  
APPENDIX 
137 
  
Dissipation 
factor (D)
+
 
in ppm 
1 ppm ~ 9999 ppm 
 
 
Quality 
factor (Q)** 
in ppm 
1 ppm ~ 9999 ppm 
  
DELTA 
0.00001% ~ 99999% 
 
*s=series, p=parallel ESR=Rs  
** with L or R  
+ with C 
Note: Only LCR-821 has Z/θ and L/R measurement 
parameters. 
 
If any of these quantities is negative, the “-“ negative 
indicator is displayed. 
Accuracy 
LCR-821/819/817 
 
 
 
R, L, C, Z 
0.05%(Basic) 
 
 
D, Q 
0.0005(Basic) 
 
 
θ 
0.03˚(Basic) 
 
LCR-829/827/826/816 
 
 
R, L, C, Z 
0.10%(Basic) 
 
 
D, Q 
0.001 
 
 
Please refer to the accuracy definition 
on page132 for details. 
Basic Accuracy 
0.05% 
LCR-821/819/817 
 0.1% 
LCR-829/827/826/816 
Test Frequency 
LCR-821 
12Hz~200kHz (504 Steps) 
 
LCR-819/829  12Hz~100kHz (503 Steps) 
 
LCR-817/827  12Hz~10kHz (489 Steps) 
 
LCR-816/826  100Hz~2kHz (245 Steps) 
Measurement 
displays 
Value 
R/Q, C/D, C/R, L/Q, Z/ θ, L/R 
*The resolution of primary display (L, 
C, R or Z) is five digits. 
*The resolution of secondary display 
(D, Q, R with C, or R with 
L) is four digits. 
*The resolution of secondary display 
(θ) is 2 digits after decimal place.
   LCR-800 
User 
Manual 
138
 
 
Delta% 
DELTA% shows the percent deviation 
of the measured L, 
C, R or Z value from a saved 
NOMINAL VALUE. The deviation is 
indicated.
 
Delta 
Delta is similar to the DELTA% except 
that the deviation is shown in suitable 
units (ohms, henries, etc.) 
Measurement 
Speed 
Speed 
LCR-
816/817/819/821 
LCR-826/827/829 
 
Slow 
896ms 
Please refer to the 
Handler timing 
diagram on page 
119 for details. 
 
Medium 
286ms 
 Fast 
135ms 
Equivalent circuit  Parallel  
L/R, L/Q, C/D, C/R, R/Q 
Serial 
L/R, L/Q, C/D, C/R, R/Q, Z/θ 
Trigger Auto/Manual 
 
Average 
1-255 
 
Battery 
3V-DC lithium ion (*BR-2/3A) used for memory and 
calibration data backup. (Recommended replacement 
every three years. 
*The battery should only be replaced by a GW Instek 
approved service center. 
Memory 
100 blocks of memory 
 
Display 
240X128 dot matrix C.C.F.L. back lit LCD (contrast 
adjustable) 
Test voltage 
LCR-817/819/827/829/821  LCR-816/826 
 
5mV~1.275V (5mV steps)  0.1V~1.275V (5mV steps) 
 
Note: When the test frequency is at 200kHz, test 
voltage must be greater than 100mV. 
DC bias 
Internal 
2V 
 
External 
Up to 30VDC (200mA max), tolerable 
up to 35VDC.   
Operation 
Environment 
Indoor use 
Altitude up to 2000M 
Installation category II 
Pollution degree 2 
Operating temperature 10°C~50°C, <85% relative 
humidity 
Storage 
Environment 
-20°C~60°C