Keysight Technologies USB data capturing module U2331A Input: 3 MS/s, output: 1 MS/s U2331A OPT 902 + ABD Manuale Utente

Codici prodotto
U2331A OPT 902 + ABD
Pagina di 71
10
U2300A Series DAQ Service Guide
1
Characteristics and Specifications
Electrical Measurement Specifications
Basic Multifunction USB DAQ Device
Analog Input Measurement
[1]
Model Number
U2351A | U2352A
U2353A | U2354A
Function
23 
°C ± 5 °C
°C to 18 °C
28 
°C to 45 °C
23
°C ± 5°C
0
°C to 18°C
28
°C to 45°C
Offset Error
±1 mV
±5 mV
±1 mV
±5 mV
Gain Error
±2 mV
±5 mV
±2 mV
±5 mV
–3 dB small signal bandwidth
[2]
760 kHz
1.5 MHz
1% THD large signal bandwidth
[2]
300 kHz
300 kHz
System noise
1 mVrms
2 mVrms
1 mVrms
2.5 mVrms
CMRR
62 dB
62 dB
Spurious-free dynamic range (SFDR)
[3]
88 dB
82 dB
Signal-to-noise and distortion ratio (SINAD)
[3]
80 dB
78 dB
Total harmonic distortion (THD)
[3]
–90 dB
–82 dB
Signal-to-noise ratio (SNR)
[3]
80 dB
78 dB
Effective number of bits (ENOB)
[3]
13
12.6
Analog Output Measurement
[1]
Model Number
U2351A | U2353A
Function
23 
°C ± 5 °C
°C to 18 °C
28 
°C to 45 °C
Offset Error
±1 mV
±4 mV
Gain Error
±4 mV
±5 mV
Slew rate
19 V/µs
Rise time
0.9 µs
Fall time
0.9 µs
Settling time to 1% output error
 4 µs
Driving capability
5 mA
Glitch energy
5 ns-V (Typical), 80 ns-V (Maximum)