Keysight Technologies USB data capturing module U2331A Input: 3 MS/s, output: 1 MS/s U2331A OPT 902 + ABD Manuale Utente
Codici prodotto
U2331A OPT 902 + ABD
Characteristics and Specifications
1
U2300A Series DAQ Service Guide
11
[1] Specifications are for 20 minutes of warm-up time, calibration temperature at 23
°C and
input range of ±10 V.
[2] Specifications are based on the following test conditions
.
[3] Specifications are based on the following test conditions
.
Bandwidth Test
Model Number
Test Conditions
(DUT setting at ±10 V bipolar)
(DUT setting at ±10 V bipolar)
•
–3 dB small
signal bandwidth
•
1% THD large
signal bandwidth
signal bandwidth
U2351A
U2352A
U2352A
Sampling Rate:
Input voltage:
Input voltage:
•
–3 dB small signal bandwidth
•
1% THD large signal
bandwidth
bandwidth
250 kSa/s
10% FSR
FSR –1 dB FS
FSR –1 dB FS
U2353A
U2354A
U2354A
Sampling Rate:
Input voltage:
Input voltage:
•
–3 dB small signal bandwidth
•
1% THD large signal
bandwidth
bandwidth
500 kSa/s
10% FSR
FSR –1 dB FS
FSR –1 dB FS
Dynamic Range Test
Model Number
Test Conditions
(DUT setting at ±10 V bipolar)
(DUT setting at ±10 V bipolar)
SFDR, THD, SINAD,
SNR, ENOB
SNR, ENOB
U2351A
U2352A
U2352A
Sampling Rate:
Fundamental Frequency:
Number of points:
Fundamental input voltage:
Fundamental Frequency:
Number of points:
Fundamental input voltage:
250 kSa/s
2.4109 kHz
8192
FSR –1 dB FS
2.4109 kHz
8192
FSR –1 dB FS
U2353A
U2354A
U2354A
Sampling Rate:
Fundamental Frequency:
Number of points:
Fundamental input voltage:
Fundamental Frequency:
Number of points:
Fundamental input voltage:
500 kSa/s
4.974 kHz
16384
FSR –1 dB FS
4.974 kHz
16384
FSR –1 dB FS