Calibration
2
U2300A Series DAQ Service Guide
35
Step 1: DAQ self- calibration
Step 1:
Power on DAQ device and disconnect all connections from DAQ.
Warm up for 30 minutes.
Power on DAQ device and disconnect all connections from DAQ.
Warm up for 30 minutes.
Step 3: Clear DAQ
Send “*CLS”
Step 4:
Wait: 1 second
Step 2: Reset DAQ
Send “*RST”
Step 5: Start self-calibration
Send “CAL:BEG”
Wait: 60 seconds
Step 6: Check operation complete status
Send “*OPC?”
OPC status = 1?
Step 7: Read system error
“SYST:ERR?”
System error?
Calibration DONE!
Yes
No
Calibration FAIL!
Yes
No