Keysight Technologies USB data capturing module U2331A Input: 3 MS/s, output: 1 MS/s U2331A OPT 902 + ABD Manuale Utente

Codici prodotto
U2331A OPT 902 + ABD
Pagina di 71
42
U2300A Series DAQ Service Guide
3
Dismantle Procedures
General Disassemble
This chapter provides the step- by- step guides on how to 
dismantle the module and install the replacement assembly. To 
assemble back the module, follow the instructions in reverse 
order. 
 
The removable assemblies include:
• Plastic casing
• Metal casing
• Rear metal casing
• Front metal casing, which is attached to the carrier board 
and measurement board
Mechanical Disassemble
Follow the instructions in this section for the instrument 
disassemble process.
Step 1: Pull the bumper out to remove the plastic casing.
N O T E
The parts shown in the following figures are representative and 
may look different than what you have in your module.
Plastic casing