Gw Instek PEL-3211 Electronic Load PEL-3211 Scheda Tecnica
Codici prodotto
PEL-3211
PEL-3000 User Manual
82
Display
2.596
V
500.06
mA
1.2982
w
CV B Value
15.000
CV A Value
15.000
LOAD
RS232
A Value
Fine
03/Sep/2012
V
V
OPP Alarm
Please Press Enter
To Clear Alarm
OPP
OCP indicator
Alarm message
when OPP is set
to Load Off
when OPP is set
to Load Off
Mode
CV
I Range
L 0.7A
V Range
L 15V
Response
Fast
Configure
UVP
Description
If the UVP is tripped, the PEL-3000 will turn off
the load.
the load.
The UVP levels can be set from 0V to 10%
higher than the rating voltage.
higher than the rating voltage.
Operation
1. Press
Main
> Configure[F5] > Protection[F1]
and set the UVP Level.
Range:
UVP Level: OFF, 0~Rating voltage +
10%
10%
Alarm
The UVP indicator and a message will only
appear on the screen when the input voltage is
below the UVP level. The Enter key must be
pressed to clear the alarm message.
appear on the screen when the input voltage is
below the UVP level. The Enter key must be
pressed to clear the alarm message.
To clear the UVP indicator, remove the cause of
the under voltage ~ i.e., increase the input
voltage.
the under voltage ~ i.e., increase the input
voltage.