338 VT1422A Command Reference
[SENSe]
[SENSe:]DATA:FIFO:RESet
[SENSe:]DATA:FIFO:RESet
clears the FIFO of values. The FIFO counter is reset
to 0.
Comments
•
When Accepted: Not while INITiated
•
Related Commands: SENSE:DATA:FIFO
…
•
*RST Condition: SENSE:DATA:FIFO:RESET
•
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
SENSE:DATA:FIFO:RESET
Clear the FIFO
[SENSe:]FREQuency:APERture
[SENSe:]FREQuency:APERture <gate_time>,<ch_list>
sets the gate time for
frequency measurement. The gate time is the time period that the SCP will allow for
counting signal transitions in order to calculate frequency.
counting signal transitions in order to calculate frequency.
Parameters
Comments
•
If the channels specified are on an SCP that doesn't support this function,
an error will be generated. See the SCP's User's Manual for its capabilities.
•
Related Commands: SENSe:FUNCtion:FREQuency
•
*RST Condition: 0.001 s
•
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
SENS:FREQ:APER .01,(@144)
set channel 44 aperture to 10 ms
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
gate_time
numeric (float32)
0.001 to 1 (0.001 resolution)
seconds
ch_list
string
100 - 163
none