338 VT1422A Command Reference  
[SENSe]
[SENSe:]DATA:FIFO:RESet
[SENSe:]DATA:FIFO:RESet 
clears the FIFO of values. The FIFO counter is reset 
to 0.
Comments
When Accepted: Not while INITiated
Related Commands: SENSE:DATA:FIFO
… 
*RST Condition: SENSE:DATA:FIFO:RESET
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
SENSE:DATA:FIFO:RESET
Clear the FIFO 
[SENSe:]FREQuency:APERture
[SENSe:]FREQuency:APERture <gate_time>,<ch_list
sets the gate time for 
frequency measurement. The gate time is the time period that the SCP will allow for 
counting signal transitions in order to calculate frequency.
Parameters
Comments
If the channels specified are on an SCP that doesn't support this function, 
an error will be generated. See the SCP's User's Manual for its capabilities.
Related Commands: SENSe:FUNCtion:FREQuency
*RST Condition: 0.001 s
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
SENS:FREQ:APER .01,(@144)
set channel 44 aperture to 10 ms 
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
gate_time
numeric (float32)
0.001 to 1 (0.001 resolution)
seconds
ch_list
string
100 - 163
none