Blender
Air-Oxygen
WARNiNg
An Oxygen Analyzer/monitor must be used to verify 
oxygen concentration.
Accuracy of oxygen concentration will be affected if bleed 
is not activated at flow settings below 15 lpm for the High 
Flow Blender, and 3 lpm for the Low Flow Blender.
dO NOT obstruct the alarm.
dO NOT use Blender when alarm is sounding.
dO NOT use oil in or around the Blender.
dO NOT occlude or obstruct the bleed port on the 
auxiliary outlet of the Blender.
dO NOT use near any type of flame or flammable/
explosive substances, vapors or atmosphere.
Oxygen Concentration dial does not rotate 360 degrees. 
Rotating the dial less than 21% or over 100% oxygen will 
damage the Blender.



CAUTiON
Turn off gas supplies when Air-Oxygen Blender is not in use.
Store the Air-Oxygen Blender in a clean, dry area when 
not in use.
The Air-Oxygen Blender contains magnetic, ferrous 
material that may affect the results of an mRI.
Ensure all connections are tight and leak free.
Avoid excessive pressure surges greater than 100 psi 
(6.9 bar) when pressuring the Blender inlets.
dO NOT steam autoclave.
dO NOT immerse Air-Oxygen Blender into any liquid.
dO NOT gas sterilize with (ETO) Ethylene Trioxide.
dO NOT use if dirt or contaminants are present on or 
around the Blender or connecting devices.
dO NOT smoke in an area where oxygen is being 
administered.
dO NOT clean with aromatic hydrocarbons.