Precision Trading PM5300 Manuale Utente

Pagina di 16
5
Blender
Air-Oxygen
SPECiFiCATiONS 
Model
PM5200 High Flow
PM5300 Low Flow
Primary Outlet
Flow Range
15 - 120 lpm
3 - 30 lpm
With both supply pressures at 50 psi (3.4 bar) 
with NO BLEED
Auxiliary Outlet
Flow Range
2 - 100 lpm
0 - 30 lpm
With both supply pressures at 50 psi (3.4 bar) 
with BLEED
Bleed Flow
13 lpm or less
at 50 psi (3.4 bar)
3 lpm or less
at 50 psi (3.4 bar)
Maximum 
Combined
Flow (All Outlets)
≥ 120 lpm
≥ 30 lpm
Bypass Flow
(Loss of Air or 
Oxygen supply)
> 85 lpm
> 45 lpm
Bypass Alarm 
Activation
50 psi 
(3.45 bar)
60 psi
(4.14 bar)
50 psi
(3.45 bar)
60 psi
(4.14 bar)
13-25 psi 16-24 psi 18-22 psi 16-24 psi
0.9-1.7 bar 1.1-1.65 bar 1.2-1.5 bar 1.1-1.65 bar
Alarm Reset:
When pressure differential is 
6 psi (0.4 bar) or less.
Alarm Sound Level:
≥ to 80 db at 1 ft (0.3 m)
Oxygen Concentration 
Adjustment Range:
21 - 100%
gas Supply Pressure: 30 - 75 psi (2.1 - 5.2 bar) 
Air and Oxygen within 10 psi (0.67 bar) 
of each other
Mixed gas Stability:
±1% Oxygen
Connection Types:
DISS Type - Air & Oxygen Inlets & Outlets 
and / or 
NIST Type - Air & Oxygen Inlets