Omega Engineering DP3300 Series Manuale Utente

Pagina di 28
DP3300 Series User’s Manual 
Page15 
1.  Connect a thermocouple calibration source to the unit. 
2.  Dial in 1100 degrees centigrade on the calibrator (Note: unit must be programmed for displaying in centigrade). 
3.  Adjust the gain pot (see diagram of rear view of instrument) on the back of the instrument until the display reads '1100'. 
4.  Short Channel #1's input with a wire or shorting bar (make sure to disconnect the calibrator so as not to damage it). 
5.  Push  DATA  key. The display will read '0002' --- or some other value. 
6.  Adjust offset pot on back of  instrument (ref. Fig. 3) until the display reads '0000'. 
7.  Push PROG key once -- the display will show Cold Junction temperature. 
8.  Remove the shorting bar from Channel one input and connect the thermocouple calibrator again. 
9.  Repeat steps 2 through 7 till the unit reads proper temperature.  Press ‘PROG’ key to step thedisplay to next Channel. 
 
 
Input Range Setup (for current and voltage): 
 
Voltage Range Setup (0-5vdc, 0-10vdc) 
 
 After calibration, the VOLTAGE range may be setup.  The display will briefly show 'U   rAnGE'  (for "Voltage Range"). Then 
the display will show the present setting.  Use ^v key (ref. Setup For High Scale) to set the desired VOLTAGE value.  Once the 
desired VOLTAGE value is displayed, push SETUP key to enter that value and go to the next step. 
 
Voltage Range Setup (for millivolt inputs) 
 
For Millivolt input units (i.e. the units that have millivolt input option instead of thermocouples), the display will first show  'HV 
rnGE'
  (for "High Voltage Range").  After the High Voltage range is set,  the display shows  'LV rnGE'  (for "Low Voltage 
Range" i.e. for millivolt input). Enter values for respective inputs. 
 
Current Range Setup 
 
After the VOLTAGE range, the CURRENT range may be set.  The display will briefly show 'C   RANGE'  (for "Current 
Range") followed by previous value.  Use ^v key (ref. Setup For High Scale) to set the desired CURRENT value.  Once the 
correct CURRENT range is displayed, push SETUP key to enter that value and return to 'SYS   CH' display. 
 
Correct Range Settings 
 
 
Following RANGE values should be entered for various inputs: 
 
 
INPUT   
 
 
RANGE Value 
  0-5 
Vdc 
    
5.000 
  0-10 
Vdc    
10.000 
  0-100 
Mv    
100.0 
 
 
4-20 Ma. (loop current) 
 
 
20.00 
 
Setup Examples 
 
Example #1: Use of the OFFSET Parameter 
 
Setup for a 4-20ma transducer signal corresponding to 0-500 gallons of fluid (reading in 10th of a gallon resolution): 
 
  DECIMAL 
POINT 
 = 
999.9 
 
 OFFSET   = 
4.000 
  HIGH 
SCALE 
  = 
500.0 
 
 TARE 
   = 
0.000 
  LOW 
SCALE 
  = 
000.0 
 
Example #2: Use of the LOW SCALE Parameter 
Setup for a 4-20ma transducer signal corresponding to 500-2000 degrees Fahrenheit temperature i.e. 500 degrees at 4ma and 2000 
degrees at 20ma (one degree resolution): 
 
  DECIMAL 
POINT 
 = 
9999 
 OFFSET   = 
4.00 
  HIGH 
SCALE 
  = 
2000 
 TARE 
   = 
0.000 
  LOW 
SCALE 
  = 
500