SETUP FOR A PULSE CHANNEL: 
 
Turning a Channel ON or OFF 
 
The very first thing after selecting a channel is to turn it on or off. A channel should be turned off  if it is not being used. Use ^v 
key to select the desired setting.  If a channel is OFF, the display will show  'CHx  OFF'  (where ‘x’ is the channel number) and 
if it is ON the display is 'CHx  ON'. Use ^v key to display the desired ON/OFF setting, and then push SETUP key to go to the 
next parameter. 
 
MODES OF OPERATION FOR A PULSE INPUT CHANNEL: 
 
A pulse Channel can operate in any one of the following modes: 
 
 
      MODE 
 
       DISPLAY SHOWS   
 
 
DESCRIPTION 
 
 
 Counter 
Mode 
  COUntEr      
    
For a Channel being configured as a counter. 
 Rate 
Mode 
  rAtE            
 
For a Channel being configured as a rate monitor. 
 RPM 
mode 
  SyS  rP         
 
For a Channel being configured as RPM monitor. 
 
High Frequency Mode 
 
HI FrEq      
    
For a Channel being configured as high frequency meter. 
 
Low Frequency Mode 
 
LO FrEq      
    
For a Channel being configured as low frequency meter. 
 
To change operating mode of the system push ^v key. After obtaining the desired mode, push SETUP key. 
 
The next function to be configured depends on the operating mode selected above. To configure for different modes, look under  
appropriate headings.  
 
 
SETUP FOR COUNTER: 
 
If a channel is setup as a counter, then the following steps have to performed. To abort at any point during the setup, push RESET 
key once to get to the very beginning of SETUP, or push and hold (till SAVING is displayed) to get out of SETUP altogether. 
 
FUNCTION
DISPLAY SHOWS
TO CHANGE
Counter  mode  0  
 
Cntr  0         
 
 
 
Push ^v key till   
 
Counter  mode  1   
     
Cntr  1     
     
 
 
desired mode  is   
Counter  mode  2  
 
Cntr  2    
 
 
 
selected . 
Counter  mode  3  
 
Cntr  3   
 
      
Counter  mode  4  
 
Cntr  4   
 
 
Counter  mode  5     
 
Cntr  5    
 
 
 
 
Preset value entry  
 
'PrESEt' followed by preset  
 
Push ^v key to increment or decrement 
                     
 
 
value e.g. 10.024 with  
 
 
active digit. Push <NXT> key to change 
                     
 
 
active digit flashing. 
 
 
active digit. 
  
Relay Delay entry 
 
'dELAY' followed by preset  
 
Push ^v key to increment or decrement 
                     
 
 
value e.g. 24 with active   
 
active digit. Push <NXT> key to change 
                     
 
 
digit flashing. 
 
 
 
active digit. 
 
Decimal point        
    
dP  9999          
 
 
 
Push ^v key to 
placement             
 
dP  999.9         
 
 
 
move decimal 
                      
 
 
dP  99.99         
 
 
 
point to desired 
                     
 
 
dP   9.999        
 
 
 
position. 
 
Scaler               
 
 
'SCALER' followed by preset  
 
Push ^v key to increment or decrement 
                     
 
 
Scaler value e.g. 0.00024 with  
 
active digit. Push <NXT> key to change