Standard Operating Parameters 
Material Present Input 
Mode 
Controls how the gauge determines Material Present.  The sensor can use 
either the amount of reflected laser light (Internal), or it can use an external 
input signal (Active Low/Active High). 
Material Present 
Dropout Time 
Sets the time that Material Present can be lost before the sensor triggers the 
end of a measurement.  If you are taking measurements and the Material 
Present signal is lost for less than the selected Material Present dropout time, 
the loss is considered to be a momentary loss of signal and is ignored. If the 
signal is lost for longer than the selected Material Present dropout time, it is 
considered the end of the material. This function is very useful when making 
length measurements on poor reflecting material. 
Material Present 
Validation Time Limit 
Sets the maximum time that can elapse between the detection of Material 
Present and the acquisition of valid velocity readings.  If the acquisition of valid 
velocity readings exceeds this time limit the gauge will consider the length 
measurement to be invalid and output 999999.999 as a length. 
Material Present 
Threshold 
A value from 0 to 4095 that indicates the relative reflected laser light threshold 
level of the Material Present signal.  A setting of 0 would take very little 
reflected light to trigger Material Present.  A setting of 4096 will take a lot of 
light on the detector to trigger MP.  A good starting place is in the center of the 
range.  This setting is only applicable when the Material Present Input Mode is 
set to Internal. 
Walking Threshold 
The velocity value below which, the output is forced to zero regardless of 
whether or not valid readings are obtained. This setting is useful for eliminating 
accumulation of length when the material is vibrating while stationary in the 
measurement zone.  Setting this value to 0 disables the setting. The range is 0 
to 2000 
Minimum Final Length 
When in Final Length mode, sets the minimum length that will be reported.  
This is useful for ignoring momentary glitches in Material Present caused by 
outside disturbances such as water droplets or spray passing through the 
measurement area.  Any measured length less than this setting will not be 
reported as a Final Length.  The range is 0 to 999,999.999. 
Index Pulse Length 
Sets the Index pulse length in 10 ms increments.  The range is from 10ms to 2 
seconds. 
Index Pulse Every 
Sets the distance between index pulses in the selected measurement units.  
The range is 0.1 and 2000 
High Speed Pulse Rate 
Sets the number of pulses per unit length for the high speed pulse output.  The 
unit length can be configured by setting the Units of Measurement setting.  The 
gauge is capable of transmitting pulse rates as high as 5MHz on this pulse 
output.  Select a Pulse Rate setting that will produce a frequency lower than 
5MHz at the maximum expected line speed.  The range is 0.001 to 100,000. 
Low Speed Pulse Rate 
Sets the number of pulses per unit length for the adjustable voltage pulse 
output.  The unit length can be configured by setting the Units of Measurement 
setting.  The sensor is capable of transmitting pulse rates as high as 250kHz 
on this pulse output.  Select a Pulse Rate setting that will produce a pulse rate 
lower than 250kHz at the maximum expected line speed. 
For example, if the maximum line speed was 1000 m/min, a setting of 10,000 
pulses/meter would produce an output frequency of about 167kHz.  (1000