Appendix E: Troubleshooting Guide 
 
 
Part No. 93463 / Drawing No. 0921-01561 
Page 217 of 221 
Revision A (Sep 2007) 
Pulses look funny on an 
oscilloscope. 
Wiring 
The pulse outputs should look 
something like the picture: 
 
 
In this picture: 
User V
IN
 = +24V
DC
 
Channel 1: Phase A True 
Channel 2: Phase A False 
Channel 3: Phase B True 
Channel 4: Phase B False 
 
If the pulses don't look like this, then the 
oscilloscope may not be connected 
properly, or the pulses/counter may be 
wired incorrectly. 
•  Check that you are connected to the 
correct pins 
•  Check that the scope ground leads 
are connected to Signal Ground, 
and not Phase A False and Phase B 
False. 
•  Check that Phase A False and 
Phase B False are not connected to 
signal ground. 
 
 
 
Miscellaneous 
 
Gauge resets itself, (Laser On 
flashing, shutter opening/closing) 
Power Supply voltage is too 
low on a long cable 
Check +24VDC supply.  Adjust up to 
+29VDC if necessary. 
 
See Appendix D for more information 
 
Laser is flashing 
Ambient temperature is too 
high (or low.) 
If the ambient temperature is too high, 
the gauge may not be able to properly 
control the laser temperature. 
 
Connect air or water coolant to reduce 
the gauge temperature.  See the 
Specifications
 section for details. 
 
System Status Indicator is OFF 
Laser At Temperature 
 
 
 
 
Laser Interlock 
If in a hot environment, cooling may be 
required for the gauge to maintain the 
proper laser temperature.  See the 
specifications section for details. 
 
The Laser Interlock (Pin 16) must be