Laser beta lasermike LS8000-3 Manuale Utente

Pagina di 221
LaserSpeed 8000-3 Instruction Handbook
 
Interfacing with the LS8000-3 
 
 
Part No. 93463 / Drawing No. 0921-01561 
Page 51 of 221 
Revision A (Sep 2007) 
Measurement Direction Input (Pin 14)  
This is an optically isolated input that can be used to externally control the 
Measurement Direction feature of the LS8000-3.  The input requires a voltage 
high (5 to 24V) to activate.  The input level should be the same as the voltage 
connected to the User V
IN
 input (Pin 20) or 5V if there is nothing connected to 
the User V
IN
 input. 
 
 
Optically Isolated Input Schematic 
 
There are two ways to reverse the direction of the LS8000-3's measurements: 
1.  Modify the Direction Inversion setting with LaserTrak (or using the "A" 
serial command) 
2.  Connect 5-24V to the Measurement Direction Input 
 
Changing the Direction Inversion setting with LaserTrak to invert the direction has 
the same effect as changing the state of this input. 
Note:    If the Direction Inversion setting is set to "Reverse +/- Directions" and 
the Measurement Direction Input is high, they will cancel each other 
out and the LS8000-3 will operate in the default manner. 
Direction 
Inversion Setting  Measurement Direction Input  LS8000-3 Direction 
Normal 
Low (Disconnected or 0v) 
Normal 
Normal High 
(5-24V) Reversed 
Reversed 
Low (Disconnected or 0v) 
Reversed 
Reversed High 
(5-24V)  Normal