Laser Interlock (Pin 16)  
The Laser Interlock input will switch off the laser power when the connection to 
ground is broken. This is commonly used with external switches for laser 
safety.  When the interlock pin is connected to ground, the laser is operational. 
When the interlock circuit is opened (break the connection to ground), power 
to the laser is switched off. The power to the rest of the LS8000-3 is not 
affected. 
Note:   To enable laser operation, connect the laser interlock pin (Pin 16) to 
ground (Pin 11, 19, or 21).  There is a delay of several seconds from 
the time the Laser On light turns on to the time that laser radiation is 
emitted. 
Signal Ground
(Pin 11, 19, 21)
Laser Interlock (Pin 16)
Safety Interlock Switch
(Close to Turn On Laser)
+5V
GND
 
Laser Interlock Input Circuit 
 
Panel Cutout
for Keyswitch
0.42in
[10.7mm]
0.50in
[12.7mm]
Laser Off
Laser On
 
Laser Interlock Key switch Connection