Hach sc1000 Manuale Utente

Pagina di 150
34
Installation
3.6.4 Modbus card connections
Modbus RS485 (YAB021) and Modbus RS232 (YAB047) are available. For more detailed 
information refer to the bus system manual.
To make a Modbus card connection:
1. Remove power from the instrument. Remove the probe module cover.
2. Connect the Modbus card to the appropriate slot (
). Use a magnetic 
screwdriver to secure the four screws to the card. 
3. Install the card connector to the appropriate connection on the main circuit board 
(
4. Feed the cable through the base of the module and properly prepare and insert each 
wire into the terminal according to 
/
 an
5. Write the serial number from the rating plate on the supplied sticker and attach it to the 
main high voltage barrier (
). 
6. Install the probe module cover. 
After installation and connection of a plug-in expansion card, the card must be configured 
to the system. For Modbus card setup instructions, refer to 
)
Figure 24 Modbus RS485 (YAB021) card connections
1
Card (Reverse side)
3
Jumper 1&2 unplugged for full duplex (4-wire)
2
Jumper 1&2 plugged for half duplex (2-wire)
4
Terminal block (Refer to 
 for terminal 
assignments)
Table 7  Modbus RS485 card (YAB021) terminal assignments
Terminal 
Modbus RS485 designation with 4 wires
Modbus RS485 designation with 2 wires
1
Not Used
Not Used
2
Not Used
Not Used
3
Output – 
 –
4
Output + 
+
5
Input –
 –
6
Input + 
+
7
Shield (connected to Protective Earth)
Shield (connected to Protective Earth)