Hach sc1000 ユーザーズマニュアル

ページ / 150
34
Installation
3.6.4 Modbus card connections
Modbus RS485 (YAB021) and Modbus RS232 (YAB047) are available. For more detailed 
information refer to the bus system manual.
To make a Modbus card connection:
1. Remove power from the instrument. Remove the probe module cover.
2. Connect the Modbus card to the appropriate slot (
). Use a magnetic 
screwdriver to secure the four screws to the card. 
3. Install the card connector to the appropriate connection on the main circuit board 
(
4. Feed the cable through the base of the module and properly prepare and insert each 
wire into the terminal according to 
/
 an
5. Write the serial number from the rating plate on the supplied sticker and attach it to the 
main high voltage barrier (
). 
6. Install the probe module cover. 
After installation and connection of a plug-in expansion card, the card must be configured 
to the system. For Modbus card setup instructions, refer to 
)
Figure 24 Modbus RS485 (YAB021) card connections
1
Card (Reverse side)
3
Jumper 1&2 unplugged for full duplex (4-wire)
2
Jumper 1&2 plugged for half duplex (2-wire)
4
Terminal block (Refer to 
 for terminal 
assignments)
Table 7  Modbus RS485 card (YAB021) terminal assignments
Terminal 
Modbus RS485 designation with 4 wires
Modbus RS485 designation with 2 wires
1
Not Used
Not Used
2
Not Used
Not Used
3
Output – 
 –
4
Output + 
+
5
Input –
 –
6
Input + 
+
7
Shield (connected to Protective Earth)