Eclipse Combustion EGA4 Manuale Utente

Pagina di 50
 
 
 
Instruction Manual MM850481 ed.01 
1.1 Specifications 
 
Type: 2, 3 or 4 cells palm-top flue gas analyzer. 
 
Calibration: automatic calibration procedure at instrument switching-on. 
 
Self-diagnosis: Sensors efficiency test with anomalous status announcement. 
 
Fuel types: Up to 10 totally programmable. 
 
Pump: rate of flow 0.8 lit / head –230mbar 
All data measured using 3 meters long probe and line filter connected. Load loss = 10 mbar using a 3 meters probe 
extension 
 
Power supply: High capacity Ni-MH rechargeable battery pack / external battery charger. 
 
Charging time : 8h at 90% with instrument off. 
 
Battery life: 5h continuous operation (without printer and back-light). 
 
Printer: Impact type 24 columns with 58 mm large and 18 meters long paper roll  
 
Printer power supply: using the analyzer battery pack. 
 
Print autonomy: up to 30 reports with battery full 
 
Automatic CO(ppm) and CO(%) selector: if the sensors are installed. 
 
Internal data memory: up to 250 complete analysis data structured by Tags. 
 
Service and user data: 3 programmable lines for each programmed customer using a PC and DBGas Software. 
 
Report Herder: 4 rows x 16 characters programmable from keyboard 
 
Display: Large (40x56 mm) graphic LCD display with automatic back-light device. Bargraph and trend capability as 
option. 
 
Serial interface: Bi-directional infrared interface. External IR/RS232 converter available for PC communication. 
 
Synchronous serial port: for external keyboard and optical bar pen reader. 
 
Smoke measurement: Using the optional external manual pump. Bacharach index memory store and printout 
capability as standard. 
 
Optional probes: ambient CO, Bar pen, explosive gas leakage sniffer, T+RH% probe. 
 
Working temperature: from –5°C to +45 °C (up to 50°C for short time) 
 
Storage temperature: from –20 to +60°C (3 months maximum at temperatures exceeding the operational limits). 
 
Dimensions: 115x90x330 mm 
 
Weight: 1.1 kg battery and printer included 
 
 
External optional ambient CO probe 
 
Range: from 0 to 500ppm 
 
Accuracy: 
±5ppm up to 100ppm; ±5% up to 500ppm 
 
Resolution: 1ppm 
 
Response time: 30s (t90) 
 
Waiting starting time: 30s 
 
Working temperature: from –5°C to +45 °C 
 
 
External optional ambient T+RH probe 
 
Temperature Range: from 0°C to 85°C 
 
Temperature Accuracy: 
±1°C 
 
Temperature Resolution: 0.1°C 
 
Relative Humidity Range: from 0 to 99%RH 
 
Relative Humidity Accuracy: 
±5%RH 
 
Relative Humidity Resolution: 0.1%RH 
 
Response time: 50s (t63) 
 
Waiting starting time: 30s 
 
Working temperature: from –0°C to +50 °C (up to 85°C tip only) 
 
 
External optional natural gas leak detection probe 
 
Pre-heating time: 30s minimum 
 
Alarms indication: visual with 5 steps 
 
5 alarms levels: 100, 200, 300, 400, 500 ppm 
 
Acoustic Alarm indication: respectively 1, 2, 3, 5 beep/s, continuous 
 
Response time: 5s (t90) 
 
Alarm levels Accuracy: 
±10% at 90 days