SECTION 2.  INSTALLATION 
2.1.  Installation Considerations 
2.1.1.  Warnings and Cautions 
 
 
WARNING:  Explosion hazard! Do not mount the SZ monitor in an area that may contain 
flammable liquids, vapors or aerosols. Operation of any electrical equipment in such an 
environment constitutes a safety hazard. 
 
 
WARNING:  Shock hazard! Always disconnect AC power before working inside 
the monitor. 
 
 
CAUTION:  Drilling holes in the SZ enclosure may damage the unit and will void the 
warranty. Please use knockouts provided for electrical connections. 
 
 
CAUTION:  The SZ unit contains sensitive electronic components that can be easily 
damaged. Be careful not to touch or disturb any of these components. 
 
2.1.2.  Inspection 
The  SZ  monitor  has been thoroughly inspected and tested prior to shipment from the factory. 
Nevertheless, it is recommended that the monitor be re-checked prior to installation. Inspect the outside 
of the enclosure to make sure there are no obvious signs of shipping damage. Open the enclosure and 
inspect  the interior of the monitor  for loose components that may have become dislodged during 
shipment. If damage is discovered, please contact the nearest Bacharach Service Center for assistance. 
2.1.3.  Location of the Monitor 
Place the monitor in a location  that  will require no more than 500 ft (152.4 m) of tubing (total of sample 
line + exhaust) to reach the sampling area.  The ammonia monitor typically requires a purge air line 
(maximum of 100 feet).  The  halogen  monitor uses ambient air to zero its gas sensor,  so  the monitor 
should be located in an area that normally does not contain gas. If an exhaust line is not used, locate the 
monitor in a well ventilated area so that the exhaust gas does not accumulate around the monitor’s 
ventilation holes.  An exhaust line can be up to 100 feet in length. 
 
The  SZ  unit  should be operated in an environment that is between 32 and 122°F (0 and 50°C);  has a 
relative humidity that is from  5  to  90% non-condensing;  and is at an altitude of no more than 6,562 ft 
(2,000 m). The area should also be relatively free of dirt, grease, and oils that could adversely affect the 
operation of the monitor. 
The location should allow the monitor to be easily accessible for visual monitoring and servicing.  
2.2.  Mounting Instructions 
The SZ monitor should be installed plumb and level and securely fastened to a rigid mounting surface.  
The enclosure utilizes keyhole mounting designed for ¼  inch fasteners.  Install  the four screws as 
displayed  in the diagram  below  or by using the provided mounting template. Allow  the  screw heads  to 
protrude approximately ¼ inch.