Typical errors which may be observed with the detector preamp signal are:
Rise time is too slow (>0.5 µs). This may cause that the pile up rejector does not 
work correctly. There may be even a large amount of regular pulses rejected and 
especially the high energetic part of the spectrum seriously affected. When using 
such detectors switch pile up rejection off.
The fall time is too fast (time constant resp. 1/e fall time < 40µs). This causes that 
the pole/zero setting cannot be correctly adjusted. Consequences may be peak 
shift   and   peak   broadening   with   higher   count   rates   and   increased   low   energy 
spectrum cutoff. Try to use another preamplifier.
2.2.6 Additional external battery pack and Power Pack
Because   of   the   changed   charging   mechanism   of   the   MCA166-USB   the   older   battery 
packs   and   power   packs   cannot   be   used!   The   MCA166-USB   offers   a   higher   battery 
capacity, improved efficiency and is able to supply more power on ±24V than the older 
MCA166. So two main features offered by the battery pack / power pack are now partially 
integrated in the MCA itself and makes this devices obsolete. If more battery capacity is 
needed, please contact GBS-Elektronik GmbH.
3 Software
3.1 Overview
There are several programs to operate the MCA166-USB. All of them serve a special 
purpose. There are programs for MS-DOS and Windows (32 Bit). The MS-DOS software 
was originally developed for the HP 200LX, whose footprint coincident with the footprint of 
the   MCA166.   The   Windows   programs   were   developed   to   operate   the   MCA166   more 
comfortably.   Besides,  they  are (together  with at least  Windows 98  or Windows 2000) 
essential   to   connect   the   MCA166-USB   via   USB   cable,   for   the   rest,   only   RS232 
communication is possible.
Furthermore,   there   are   programs   and   software   components   to   view   and   evaluate   the 
measured data.
The following sections give brief descriptions of the programs. You can find more detailed 
information on the CD delivered together with the device. You should also use the online 
help of each program.
The diagnostics menu of each program shows the primary currents of the preamplifier 
power   supply   (±12V   and  ±24V).   This   are   not   the   currents   that   draws   the   detector. 
However, this currents can be easily calculated using the following formula:
10
I
out
0.8⋅I
Batt
U
Batt
U
out