2.2.3 Charging the MCA166-USB
Charging the MCAs battery is usually done by connecting the wall adapter to the ”Charger 
In” connector and power it from the mains supply. The wall adapter is able to deliver 
enough power to operate the MCA166-USB stationary under a maximum load condition 
(HV on, detector draws up to 2W preamplifier power). If the MCA166-USB is operated on 
a high power USB hub and the wall adapter isn’t connected, the MCA166-USB can be 
charged from the USB port. But there is a limitation of 2,5W nominal input power which is 
the maximum power that can be drawn from  a high power USB port (real usable input 
power depends on hub voltage, voltage loss on USB cable and connectors and efficiency 
of internal charger; it is usually about 2W). This is enough power to operate the MCA166-
USB without detector or with a low power detector (e.g. CdZnTe or NaI). If a detector 
draws to much power from the MCA166-USB (e.g. some HPGe), charging via USB is still 
working but the battery runs out slowly.
If the MCA166-USB is connected to a working high power USB hub, charging will start 
after  6 seconds  automatically.  It  is possible  to disable  USB  charging  by some  newer 
application programs such as WinSpec. This is very useful if the MCA is operated at  a 
laptop computer which runs on battery. Connecting the wall adapter to the MCA166-USB 
and supply it with power will always disable USB charging.
The  old   battery  chargers   supplied  with   the MCA166  (3  pin   LEMO   connector)   are  not 
compatible with the MCA166-USB! Don’t try to connect them to the MCA166-USB. The 
connectors are mechanically different.
2.2.4 Connecting the MCA166-USB to a Computer
Communication   between  MCA166-USB   and   host  computer   is  possible   via  a  standard 
serial connection (RS232) or via USB interface. For RS232 communication the RS232 
cable with SUB-D9 connector must be connected between the MCAs PC-socket and the 
computer, for USB communication the special USB cable with LEMO connector must be 
used respectively. Two baud rates can be used for communication on both interfaces, 
38.400 Baud and 307.200 Baud. Both baud rates are usable on both interfaces but most 
computers doesn't support 307.200 Baud on the RS232 interface. The firmware of the 
MCA166-USB   is   able   to   detect   the   baud   rate   and   changes   it   accordingly.   Newer 
application programs tries to connect the MCA with the higher baud rate first, if it doesn't 
work they try it with 38.400 Baud. This method guarantees that communication always 
works   on   the   highest   possible   baud   rate   without   user   interaction   and   maintains 
compatibility to older MCAs (supporting only 38.400 Baud). Because the firmware of the 
MCA166-USB is able to detect the baud rate, it is also possible to control this MCA with 
older application programs which support only 38.400 Baud.
Running the MCA166-USB on a Palmtop computer or Pocket PC is principal possible but 
needs a special communication cable. Please contact GBS-Elektronik for availability. For 
proper wiring see Table 5: Wiring Diagram of the PC Interface Connector (female Lemo
0S-306) 
on Page 21.