Braun MCA166-USB Manuale Utente

Pagina di 72
1 Introduction
The MiniMCA MCA166-USB is a battery powered high performance 4K Multi-Channel 
Analyzer/Multi-Channel   Scaler   module   comparable   in   its   performance   with   laboratory 
grade MCA. High voltage supply for detector and preamplifier power supply are integrated 
as well as an internal amplifier and pulse shaping network. Together with a small detector 
it forms a pocket-size gamma spectroscopy system and timer / counter, which is well 
suited to the demands of field measurements for international safeguards, environmental 
monitoring,   nuclear   waste   treatment   facilities,   radioactive   transport   control   and   similar 
applications.
Furthermore, the MCA166-USB supports a vast number of different detectors and its 4k 
resolution   is   adequate   to   support   high   resolution   gamma   spectrometry   with   HPGe 
detectors.
The MiniMCA software allows to operate the device as a general purpose multi channel 
analyzer (SPEC, WinSPEC) and multi scaler analyzer (MCS, WinMCS). Additional user 
programs   which   support   safeguards   specific   applications   as   U-235   enrichment 
verification, spectral radiation survey meter mode supporting active length determination 
are available. See chapter 3 for more information.
The device can also be operated with the ORTEC software (MAESTRO, GammaVision, 
ScintiVision etc.); contact ORTEC for details.
The firmware of the MCA166-USB (software of MCAs internal processor) is described in 
Appendix 3 and 4. Instructions how to change the firmware can be found in Appendix 5.
Basic software functions
File menu: write/read functions with drive/path and file pick list functions
Setup menu: ADC, Amplifier, Presets, Memory splitting, MCA mode, MCS mode, 
Multi spectral recording mode, automated instrument configuration using setup files 
Acquire control: Start, Stop, Clear, Presets
Automated repeated measurement functions
Display functions: Automated linear and logarithmic, manual linear Y scale, X axis, 
expansion and scrolling, Cursor function, ROI setting and processing 
Peak stabilization
Analysis functions: ROI peak area and integral, FWHM calculation
Analysis report function
Energy calibration function: calibration curve using up to 3 peaks
Incorporated help texts
5