2.2.1 Power management
The MCA166-USB works with a built in rechargeable Li-Ion battery, which has no memory 
effect  and  is deep  discharge and short circuit  protected.  If the device  is working, the 
remaining battery life time is permanently checked. When the battery voltage has dropped 
down to 6.8 Volts, while a data collection is in progress, the MCA166-USB gives out a 
warning to the user (the lamp will flash irregularly and the software will show an error 
message).   The   running   measurement   is   automatically   stopped   and   the   detector   high 
voltage together with the preamplifier power supply will be switched off.
Now the user can transfer the measured spectrum to the computer, but should switch the 
device off afterwards or connect the charger. If the user does not respond, and the battery 
life   time   is   nearly   used   up   (battery   voltage   6.5   Volts)   ,   the   device   switches   off 
automatically. It is made sure, that the set up and the spectrum gathered previously is 
saved in the memory of the MCA166-USB for at least several month.
The   remaining   battery   life   time   is   sufficient   to   transfer   the   measured   spectrum   to   a 
computer. It is recommended, before reading out the MCA166-USB, to connect it to the 
charger.
In the case that the battery voltage drops below its lowest value, the battery itself switches 
off. This can be, for example, the result of a lasting short circuit at the preamplifier power 
supply.  As  a  consequence   the   internal   setup   and   the  measured   values   are   lost.   The 
MCA166-USB has to be connected to the charger to recharge the battery.
2.2.2 Switching on
The MCA166-USB is set into operation by turning the power switch on (see Figure  1). 
After that, the green power ON LED flashes with a frequency of 2.5 Hz if no MMCA 
software runs on the connected computer. If the power switch is already in position ON 
(for example after an automatic switch off due to a complete loss of battery power) switch 
OFF and to ON again. When the power ON LED does not flash, connect the MCA166-
USB   with   the   charger   and   repeat   once   more.   The   MCA166-USB   is   on   line   with   the 
computer if the LED flashes with 1.2 Hz. Irregular flashing of the lamp indicates an error.
7