Specifications
13-5
13
Display
Display Size: 96 mm x 72 mm.
Display Type: Transflective active matrix LCD with LED backlight.
Display Resolution: 320 x 240 pixels.
Sweep Speed: 25 mm/s nominal (stationary trace; sweeping erase bar).
Viewing Time: 4 seconds.
Thermal Array Printer
Continuous Real Time Strip: User starts and stops the print strip. The Print 
Strip prints the selected ECG lead with the following data:
HEADER 1: Date, Time, Heart Rate, the SpO
2
 Value (if available), and 
the text "Delayed" if the recording has been configured for Delayed Mode. 
Prints every 12 seconds.
HEADER 2: Current mode (AED/Manual) the current Lead, the current 
Gain, filter setting, the text "Sync" if Sync has been enabled, and the Pacer 
Settings (consisting of the Pacer Mode, Rate, and Current (if currently pac-
ing the patient). Prints every 12 seconds, with Header 1.
HEADER 3: Changes in Mode, Gain, Lead, Sync, and Pacer Settings.
FOOTER: Drug Annotations, HR/SpO
2
 limits on a Limit Alarm, the 
Results of an Analysis in AED Mode (No Shock Advised, Shock Advised, 
or Cannot Analyze), Charging to xxxJ, Shock Delivered, No Shock Deliv-
ered, Disarm, Battery Low, Battery Critical.
SYMBOLS: Mark Triangle (for presses of the Mark key), an Alarm Bell 
(Alarm Limit Violations), Lightening Bolt (Shock Delivered), Vertical 
stripe Boundries/Pacer/Sync Tick Marks).
Event Printing: Mark Event key automatically documents events and ECG 
during defibrillation episodes. The Mark Event key can annotate the event 
with one of the following labels: Epinephrine, Atropine, Lidocaine, and 
Other.