Philips HEARTSTART XLT M3500B Manuale Utente

Pagina di 180
Specifications
13-6
Specifications & Safety
Thermal Array Printer (cont.)
Auto Printing: The printer can be configured to automatically print on Mark, 
Charge, Shock and Alarm.
Delayed Printing: The printer can be configured to run real time or with a six 
second delay.
Reports: The following can be printed: Event Summary, Configuration, 
Extended Self Test, System Log.
Speed: 25 mm/s with an accuracy of 
5%. 
Amplitude Accuracy: 
 10% or 
 50 uV, whichever is greater.
Paper Size: 50 mm by 30 m (100 ft).
Battery and Battery Power Module
Battery Type: 2 Ah, 12V, rechargeable, Sealed Lead Acid.
Battery Dimensions:  2.4" (H) x 0.94" (W) x 7.2" (D).
61.7 mm (H) x 23.9 mm(W) x 182 mm (D).
Weight: 1.4 lb (0.65 kg).
Charge Time: 
Approximately 14.5 hours to 100%, indicated by LED on power module.
Approximately 2.5 hours to 90%, indicated by LED on power module.
Capacity: 2.7 hours ECG monitoring or 50 full-energy discharges or 2.1 
hours ECG monitoring while pacing (with a new, fully charged battery and 
temperature above 20
o
C).
Battery Indicators: LOW BATTERY message appears on monitor when at 
least 10 minutes of monitoring time and 6 maximum-energy discharges 
remain (with a new battery and temperature above 20
o
C).
Battery Storage: Should not be stored above 40
o
C for extended periods of 
time. 
Charger Output: Unit can be operated using Class I AC power module alone 
with no battery installed.
+