Philips HEARTSTART XLT M3500B Manuale Utente

Pagina di 180
Connecting to Power
2-10
Getting Started
Removing the Battery
Before removing the battery, make sure the HeartStart XLT is powered off. 
Then, to remove the battery, press the black battery eject button and pull the 
battery out, as shown in Figure 2-7.
Figure 2-7   Removing the Battery 
Low Battery Warning
The message 
/RZ%DWWHU\ is displayed on the HeartStart XLT when the battery 
is low and needs recharging. This message indicates that the battery has suffi-
cient remaining capacity to provide only about ten minutes of monitoring time 
and six shocks before the HeartStart XLT shuts off. Replace the battery as 
soon as possible.
If the power is off for less than 2 minutes, while you change the battery, the 
HeartStart XLT assumes that you are continuing to treat the same patient. It 
continues to store data on the Data Card and append events to the existing 
Event Summary. Alarms set prior to the power loss remain active. 
If power remains off for more than 2 minutes, the HeartStart XLT assumes 
you are treating a different patient and assigns a new incident number.  A new 
Event Summary begins with the next event.