Philips HEARTSTART XLT M3500B Manuale Utente

Pagina di 180
Using a Data Card
2-11
2
Using a Data Card
Use of a Data Card is optional; the defibrillator will power up without a Data 
Card inserted. If you would like to collect patient information on a Data Card, 
the card must be inserted into the HeartStart XLT before the device is turned 
on.
CAUTION
Inserting or removing the data card while the defibrillator is on can corrupt 
the Data Card and prevent the unit from powering on again. If this occurs, see 
Table 12-3, Troubleshooting Tips.
The recommended practice is to use one Data Card per patient. Once a Data 
Card fills, recording stops; a second Data Card may not be inserted for the 
current incident, because the device will only allow the use of one Data Card 
per incident. Data Cards hold up to two hours of patient information. 
Multiple incidents can be recorded on a single Data Card. Each incident is 
assigned a unique incident number. 
Patient data from a Philips M3510A Data Card may be downloaded to a 
HeartStart Event Review Data Management system. HeartStart Event Review 
also allows you to erase patient data from a Data Card, allowing the card to be 
reused for another patient. 
It’s recommended that you use a designated Data Card to configure one or 
more defibrillators/monitors.
CAUTION
Use only the M3510A Data Card. These cards are specifically formatted to 
work with your Philips defibrillator. Generic cards, or other types of cards 
(such as modems) will not work, and may cause the defibrillator to
malfunction.