Overview
3-4
Defibrillating in AED Mode
Defibrillation 
(with a modified configuration)
Chapter 10 describes, in detail, the configurable parameters for AED Mode. 
Three parameters significantly impact the defibrillation process. They are:
Device Initiated Analysis - initiates ECG analysis when the HeartStart XLT 
is first turned on. The default configuration setting is On. If you choose to set 
this parameter to 
2II, you need to press 
 to initiate analysis in step 2 
of the defibrillation process. 
Automatic Re-analysis - initiates ECG analysis in between shocks within a 
shock series. The default configuration setting is 
2Q. If you choose to set this 
parameter to 
2II, you need to press 
 
to initiate analysis in between 
shocks within a shock series (i.e. after the first and second shock of a three 
shock series). 
Rhythm Monitoring - monitors the ECG for potentially shockable rhythms 
when the HeartStart XLT is not analyzing, defibrillating, or paused. The 
default setting is 
2QIf you choose to set this parameter to 2II, the HeartStart 
XLT will not look for potentially shockable rhythms during these idle times. 
Idle times also include:
z
power on, when Device Initiated Analysis is off
.
z
in between shocks within a shock series, when Auto Re-analysis is 
off. 
If Rhythm Monitoring is off, you need to observe the patient during idle times 
and determine when to press 
.
The following sections describe the defibrillation process in detail. They also 
describe what happens at the completion of a shock series and if a shockable 
rhythm is not detected. 