Defibrillating
3-6
Defibrillating in AED Mode
Defibrillating
Follow the voice and screen prompts as they guide you through the following 
steps: 
1. Press On. 
In this first step of the defibrillation process, the HeartStart XLT checks to 
see if the pads patient cable and the pads are connected. If they are, it pro-
ceeds to step 2. 
If the pads patient cable is not properly attached, you are prompted to 
&RQ
QHFW3DGV&DEOH.
Figure 3-3  Connect Pads Cable Display
Once the cable is connected, the HeartStart XLT checks to make sure the 
pads are making good contact with the patient’s skin. It does this by moni-
toring the electrical impedance between the two pads.
Connect
Pads
Cable
Shocks: 0