Setting the Heart Rate Alarm
4-6
Monitoring the ECG
Setting the Heart Rate Alarm
The computed heart rate (number of detected QRS complexes per minute) is 
displayed below the 
 sofkey, next to the 
. The heart rate represents 
the number of QRS complexes detected in a minute. A QRS beeper, if config-
ured to on, indicates each QRS complex detected.
If desired, a HR alarm may be set to alert you when the heart rate is outside 
the specified limits. Limit choices are listed in Table 4-4.
To set a HR alarm, cycle through the limit choices by pressing 
 until 
the desired limits are shown. The 
 then appears next to the heart rate value 
to indicate that the HR alarm is set.
WARNING
 
Heart rate alarms are temporarily suspended in AED Mode during ECG analysis or 
when 
 is pressed (for the duration of the paused period). Heart rate alarms are 
also suspended while charging for defibrillation and delivering a shock. 
WARNING
Heart rate displays and alarms function with internal and external pacemakers, but 
they can be unreliable. Observe the patient closely if pacemakers are used.
Table 4-4 HR Alarm Limit Choices
Alarm If Over:
Or under:
100
30
140
60
160
90
200
120
+5$/$50
+5$/$50
3$86(