Philips HEARTSTART XLT M3500B Manuale Utente

Pagina di 180
Troubleshooting
4-8
Monitoring the ECG
Poor ECG signal 
quality
z
The monitoring elec-
trodes are not making 
proper contact with the 
patient.
z
The monitoring elec-
trodes are outdated or 
dried-out.
z
Radio frequency inter-
ference (RFI) is causing 
artifact.
z
Check that the monitoring 
electrodes are properly 
applied. If necessary, pre-
pare the patient’s skin and 
apply new electrodes.
z
Check the date code on 
the electrodes. Do not 
open the electrode pack-
age until immediately 
prior to use.
z
Relocate or turn off equip-
ment that may be causing 
RFI.
QRS beeper inau-
dible or beeps do 
not occur with 
each QRS com-
plex.
z
The QRS beeper is con-
figured to Off.
z
The amplitude of the 
QRS complex is too 
small to detect.
z
Check that the QRS 
beeper is configured to 
On.
z
Adjust the volume.
z
Adjust the size of the 
ECG.
Table 4-5 Troubleshooting Tips (Continued)
Situation
Cause
Solution